2274

General Description

A mechanical microliter/milliliter multiple-syringe pump designed for low flow rate infusion and reduction of pulsation.

- Maximum rate using two 100 ml syringes 154.8 ml/min, slowest rate using one 1 cc tuberculin syringe .000077 cc/min
- Speed accuracy ±0.01%
- 30-speed mechanical gear box with positive locking action, range of 17,000:1, 100 RPM synchronous motor
- A gear overload clutch stops pusher block advance when the required force exceeds 25 lbs
- Syringe Support: Up to 2 syringes; glass and plastic; 10 µl to 50 µl

2017817